BWS platform scales Technical Manual
- 16 -
Options: on / off
anm
To set animal mode.
Options: on / off
P4 S T
ST ON
ST OFF
P5 CLR
CLRCAL
Clear the calibration record
CLROPT
Clear the parameter modification record
7. CALIBRATION
To set calibration, turn on the scale.
Press key during the self checking.
Display will be show pn
Press , and to enter, display will be show po chk
Press until display will be show p 2 mod.(These is a switch on the
main board you need to press it then can into the parameter)
Press key to confirm and press to select sigr /dual 1 /dual 2
Press key to confirm and press to select cal
Press key to confirm
Calibration Cal
Press key to enter calibration, display will be show unld
Remove all the weight from the platform.
When indicator get stable, press key to confirm.
Display will be show the last calibration weight. If want to change the
calibration weight value, press and keys to change the active
digits and press key to increment the value.
When the calibration value is correct, press key to confirm.