9.5.3 Controller information
9.5.4 cal-test
The cal-test function is used to test the sealing of the system measuring
head - rubber sleeve - sample bottle and the operability of the system
OxiTop
®
Control.
In the BOD Standard mode with the GLP On setting, the controller indi-
cates when the next cal test is due, depending on the set calibration
interval (see section 9.4.4 GLP (
OPERATING MODES BOD STANDARD, BOD
S
PECIAL AND PRESSURE P), page 61).
To perform the test, you need the WTW test resource, OxiTop
®
PM,
order number 209 333.
Sample preparation: see operating manual of the OxiTop
®
PM test
resource.
Query stopped!
You can select between
- Continue query (see above)
and
-Stop (return to the CHECK
menu)
CHECK menu.
Use <><> to select the -
Controller Info menu item.
The display shows the following
information:
Number of measuring heads
for which there is still memory
capacity
Status of the supply batteries
Status of the data backup
batteries (OK/LOBAT).
Use <><>
to scroll to the
specification of the
Software version
Series number
(can differ from the example
shown here)
Return to the CHECK menu:
With <RUN/ENTER>.
RUN/ENTER

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw OxiTop Control OC 110 and is the answer not in the manual?

wtw OxiTop Control OC 110 Specifications

General IconGeneral
Brandwtw
ModelOxiTop Control OC 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals