Overview OxiTop
®
Control OC 110
8 ba31116e05 07/2014
1.2 The measurement principle
The respirometric measurement is a pressure measurement. If oxygen is
consumed in a closed vessel at a constant temperature, a negative pressure
develops. If a gas is released, an overpressure develops. The OxiTop
®
measuring head measures and stores this pressure for the whole duration of a
measurement once started.
The OxiTop
®
OC 110 controller collects the pressure values from the
measuring heads and processes them.
The formula shown below is the basis for all calculations for the BOD using the
values from the OxiTop
®
measuring head.
The interpretation of the pressure differences in the temporal course depends
on the measured material and its preparation and on the sample manipulation
(e. g. intermediate aerations) during the measuring period.
1 Measuring head OxiTop
®
-C or OxiTop
®
-C/B
® ®
® ®
® ®
1


2
0
m
l
lt
m
2
Op
T
T
V
VV
TR
OM
BOD
M(O
2
) Molecular weight (32000 mg/mol)
R Gas constant (83.144 lmbar/molK)
T
0
Reference temperature (273.15 K)
T
m
Measuring temperature
V
t
Bottle volume (nominal volume in ml)
V
l
Sample volume in ml
a Bunsen absorption coefficient (0.03103)
p(O
2
) Difference of the D. O. partial pressure (mbar)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw OxiTop Control OC 110 and is the answer not in the manual?

wtw OxiTop Control OC 110 Specifications

General IconGeneral
Brandwtw
ModelOxiTop Control OC 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals