EasyManua.ls Logo

wtw pH 330i - Instrument Configuration; Configuration Procedure

wtw pH 330i
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuring pH 330i/340i
12
Configuring
Note
You can leave the configuration menu at any time with <M>. The parameters
already changed are stored.
Procedure for configuring (factory settings appear in bold typeface):
l Switch off the measuring instrument.
l While pressing the <M> key, press <>.
l Select the required Baud rate with <▲> <▼>.
Selection: 1200, 2400, 4800, 9600 Baud.
l Then press the <RUN/ENTER> key.
l Select the required display during the pH calibration with <▲> <▼>.
Selection: Buffer nominal value (/pH) or current electrode voltage (mV).
l Then press the <RUN/ENTER> key.
l Set the required calibration interval with <▲> <▼>.
Selection: 1... 7... 999 d.
l Then press the <RUN/ENTER> key.
l Select the date and time step-by-step with <▲> <▼>. In between, press
the <RUN/ENTER> key each time.
l When the last <RUN/ENTER> key has been pressed, the instrument
changes to the last active measuring mode.
Baud
0
40
8
LC
A
pi
sd
/pH
7
3n
t
pH
Time
I
d
Day.Month
4
00
9