pH 330i/340i Resetting to default settings
13
Resetting to default settings
You can reset the measurement parameters and the configuration to the de-
livery status separately from one another (initialization).
Measurement
parameters
The following measurement parameters can be reset to the delivery status:
Configuration
parameters
The following configuration parameters (InI) can be reset to the delivery sta-
tus:
Procedure of the reset:
pH/ORP (pH InI)
Measuring mode pH
Slope -59.16 mV/pH
Asymmetry 0 mV
Manual temperature input 25 °C
Resolution (pH display) 0.01
Baud rate 4800
Display during pH calibration Buffer nominal value
Interval 1
(automatically saved) OFF
Interval 2
(for data transmission) OFF
l To switch it off, press <CAL> while pressing the <RUN/ENTER> key.
l To reset the measurement parameters with <▲> <▼>, select YES and
confirm with <RUN/ENTER>, or
l Continue to the configuration parameters without resetting (InI) with the
<RUN/ENTER> key.
l After the configuration parameters, InI changes the instrument to the last
active measurement mode.