VARIO pH What to do if...
45
ba75373e03 05/2005
Display of l
Instrument does
not react
to keystroke
Sensor symbol
flashes
Message S toFull
StoFul lStoFul l
StoFul l
Adapter with electrode
+ measuring solution:
Conductivity too low Use suitable electrode
Temperature too high Use suitable electrode
Organic liquids Use suitable electrode
Cause Remedy
Battery almost empty Exchange the battery (see
section 5.2)
Cause Remedy
Operating condition undefined or
EMC load unallowed
Processor reset:
Press <1> and <8> at the
same time and release them
again.
Stored data and calibration
values are retained.
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
All memory locations are full Clear the memory (see section
4.5.2).