6-23
IM 735302-01E
Waveform Display
1
2
3
4
5
6
7
8
9
10
11
App
Index
Peak Curve Fit Target Range
Peak-curve-fits the specified trace waveform and writes the results to trace G.
Calculations are applied to mode peaks at the threshold value or more. The threshold
value is set in the range of 0 to 99 dB (steps of 1).
The trace display at the side of the data area changes to “PKCVFIT @".
Peak Curve Fit Target Trace
Soft Key Display
G= PKCVFIT A Peak-curve-fits trace A.
G= PKCVFIT B Peak-curve-fits trace B.
G= PKCVFIT C Peak-curve-fits trace C.
6.7 Curve Fitting