5-27
IM 735302-01E
Mesurement
1
2
3
4
5
6
7
8
9
10
11
App
Index
5.8 Sampling Point/Interval Settings
Procedure
The following are three ways in which the number of samples can be entered.
Setting the number of samples directly
Setting by the sampling interval
Automatically setting the optimum number of samples or the sampling interval
according to the sweep width (span) and resolution setting
1.
Press SETUP. The soft key menu for sweep condition settings appears.
2.
Press the SAMPLING POINT or SAMPLING INTERVAL soft key to set the
sampling point or sampling interval, respectively. To automatically set the sampling
points and interval according to the span and wavelength (frequency) resolution
settings, press SAMPLING POINT AUTO. The sampling points or interval setting
screen is displayed.
If you pressed
SAMPLING POINT AUTO, the sampling points and interval are set
automatically.
3.
Enter a number of samples or interval using the rotary knob, arrow keys, or
numeric key pad.
4.
Press ENTER. The sampling points or interval are set.
Automatic setting
Sets
Sets
sampling interval
sampling points
Note
The instrument displays if the settings for span, the number of sampling points, and
the resolution are inappropriate. When is displayed, normal measurement is not
possible.
See section 5.7 for the corrective actions when is displayed.