8-19
IM 735302-01E
Saving/Loading Data
1
2
3
4
5
6
7
8
9
10
11
App
Index
"HIGH 1"
Measurement sensitivity
"MEAS"
Measurement identifier
"LSUNT",0
Vertical axis scale mode (0: dBm, 1: dBm/nm)
"NMSKV","OFF"
Noise mask setting
(NMSKV:VERTICAL, NMSKH:HOLIZONTAL)
"RESCORE",0
Resolution correction setting (0: OFF, 1: ON)
"RESPARM",10372
Wavelength resolution RMS coefficient
"FREQPARM",10372
Frequency resolution RMS coefficient
The reference level and main level scale are saved to one of the following depending on
the vertical scale.
Main level scale
Vertical Axis Scale Save Format Description
LOG “REFL”,***.* Reference level
“LSCL”,***.* Level scale
Linear “REFL”,***.* Reference level
“LSCL”,***.* Level scale
“BASEL”,****.** Base level
Sub-level scale
Vertical Axis Scale Save Format Description
LOG “REFL”,***.* Reference level
“SSCLLOG”,***.* Level scale
“LOFST”,***.* Level offset
Linear “REFL”,***.* Reference level
“SSCLN
”,***.* Level scale
“SMIN”,****.** Base level
DB/km “REFL
”,***.* Reference level
“SSKM”,**.* Level scale
“OFSKM
”,***.* Offset level
“LENG
”,**.*** Optical fiber length
% “REFL
”,***.* Reference level
“SSPS”,***.* Level scale
“SMINP
”,***.* Base scale
Measurement Sensitivity
The data below is saved as measurement sensitivity depending on the measurement
sensitivity type.
Format Measurement Sensitivity Type
"NORM-HOLD" NORM/HOLD
"NORM-AUTO" NORM/AUTO
"NORMAL" NORMAL
"MID" MID
"HIGH 1" HIGH 1 (CHOP OFF)
"HIGH 2" HIGH 2 (CHOP OFF)
"HIGH 3" HIGH 3 (CHOP OFF)
"MID_SW" MID (SWITCH ON)
"HI1_SW" HIGH1 (SWITCH ON)
"HI2D_SW" HIGH2 (SWITCH ON)
"HI3D_SW" HIGH3 (SWITCH ON)
Note
If PEAK HOLD is selected in pulse light measurement mode, a P- is added to the front of the
data above. Similarly, if EXTERNAL TRIGGER is selected, an E- is added.
8.3 Saving/Loading Displayed Data