EasyManua.ls Logo

YASKAWA AQ6370B - Page 339

YASKAWA AQ6370B
383 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
App-27
IM 735302-01E
Appendix
1
2
3
4
5
6
7
8
9
10
11
App
Index
DRIFT(MEAS)
Wavelengths measured previously are used as references to display wavelength/level
changes (drifts).
Explanations of display items
NO: Channel number
REF WL: Reference wavelength of the channel
(previous wavelength measured)
MEAS WL: Center wavelength of the channel
DIFF MAX (wavelength): Maximum value of the relative wavelength to the reference
wavelength of the channel
DIFF MIN (wavelength): Minimum value of the relative wavelength to the reference
wavelength of the channel
REF LVL: Reference level of the channel (previous measurement level)
MEAS WL: Measurement level of the channel
DIFF MAX (level): Maximum value of the relative level to the reference level of
the channel
DIFF MIN (level): Minimum value of the relative level to the reference level of
the channel
Reference wavelength/level can be changed under the following conditions.
Active trace waveform data when MAX/MIN RESET is set by the parameter.
The first waveform data measured when wavelength axes (SPAN WL/START WL/
STOP WL) were changed according to the measurement conditions.
Appendix 4 Details of WDM Analysis Function