EasyManua.ls Logo

YASKAWA AQ6370B - Page 352

YASKAWA AQ6370B
383 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
App-40
IM 735302-01E
XdB stop band
Wavelength
Level [dB]
xdB
xdB stop-band
λi
nominal wavelength
Port A
Port B
Parameter
THRESH
Procedure
Obtain the width (xdB_stop-band) that centers on the reference wavelength
λ
i of each
channel and that is located below by the parameter THRESH_LEVEL both on the left
and on the right.
XdB pass-band
xdB Passband
xdB
Test Band
Minimum Level
in Test Band
λi
nominal wavelength
Wavelength
Level [dB]
Port A
Port B
Parameter
THRESH
TEST BAND
Procedure
Perform the bottom search within the range of parameter Test_Band/2 by centering on
reference wavelength
λ
i of each channel and obtain the bottom level (LBi).
Obtain the width (xdB_pass-band) that is below the bottom level LBi obtained in the
above procedure by the parameter THRESH _LEVEL.
Appendix 6 Details of Optical Filter Analysis Function