EasyManua.ls Logo

YASKAWA AQ6370B - Page 358

YASKAWA AQ6370B
383 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
App-46
IM 735302-01E
XdB_STOP-BAND
λ
i
Wavelength
Level [dB]
nominal wavelength
xdB
xdB stop-band
Port A
Port B
Parameter
THRESH
Procedure
Obtain the width (xdB_stop-Band) that centers on the nominal wavelength
λ
i of each
channel and that is located below by the parameter THRESH_LEVEL both on the left
and on the right.
XdB_ELIMINATION BAND
λ
i
Wavelength
nominal wavelength
Port A
Port B
Level [dB]
xdB Elimnation
band
xdB
Test Band
Maximum Level
in Test Band
Parameter
THRESH
TEST BAND
Procedure
Perform the peak search within the range of the parameter Test_Band/2 by centering
on the nominal wavelength
λ
i of a channel and obtain the peak level (LPi).
Obtain the width (xdB_Elimination_Wavelength) that is above the peak level LPi
obtained in the above procedure by the parameter THRESH_LEVEL both on the left
and on the right.
Appendix 6 Details of Optical Filter Analysis Function

Table of Contents