3-11
IM 735302-01E
Preparing for Measurement
Turning the Power OFF
1.
Press the POWER switch on the front panel of the instrument. A shut down
confirmation message is displayed along with the YES and NO soft keys.
2.
Press the YES soft key. The message, “AQ6370B is shutting down. Please wait...”
appears, and shut-down begins. If you do not wish to shut down, press the NO
soft key. The screen returns to the original soft key menu.
3.
After the POWER switch changes from green to orange, turn OFF the MAIN
POWER switch on the rear panel of the instrument.
CAUTION
Do not cut the power to the instrument with the MAIN POWER switch on the rear
panel when an operation is in progress. The operating system configuration file will
not be backed up, possibly resulting in malfunctions upon start up the next time the
instrument is turned ON. Always use the above procedure to shut down.
You can also shut down the instrument using panel keys and soft keys.
1.
Press SYSTEM.
2.
Press the MORE soft key three times. The SYSTEM 4/4 screen is displayed.
3.
Press the SHUT DOWN soft key.
4.
Press the YES soft key. Shut down begins.
5.
After the POWER switch changes from green to orange, turn OFF the MAIN
POWER switch on the rear panel of the instrument.
Note
If for some reason the instrument fails to shut down normally, hold down the
POWER
switch
for approximately four seconds or longer to force standby mode. Note that the operating system
configuration file will not be backed up, possibly resulting in malfunctions upon start up the next
time the instrument is turned on.
3.4 Turning the Power ON/OFF