3-18
IM 735302-01E
Using the Instrument’s Internal Calibration Wavelength Value
6.
Press the CALIB WL SELECT soft key. The soft key menu changes to show
wavelength values.
7.
Press the soft key corresponding to the appropriate wavelength value.
8.
Press the EXECUTE soft key. Wavelength calibration is executed. The previous
screen is displayed again after the calibration process ends.
9.
Press the CANCEL soft key during wavelength calibration to cancel the
wavelength calibration process.
Note
It is not necessary to perform calibration on multiple wavelengths. If calibration is performed on
multiple wavelengths, only the results from the final calibration are applied.
3.7 Wavelength Calibration