<2. Explosion Protection Type>
6
Fluid Temperature and Pressure:
The gure shows the usable range of the sensor
in each specication. It is also limited by the
ange pressure rating of the process connection.
AXG###
PFA Lining, Remote Sensor (excluding high pressure
type (ASME Class 600 ange) and Hygienic Use)
Pressure
MPa (psi)
4 (580)
2 (290)
1 (145)
– 0.1 (–14.5)
– 40
(–40)
40
(104)
100
(212)
150
(302)
Temperature °C (°F)
130
(266)
2.5 to 50 mm (0.1 to 2 in.) (flange type, wafer type)
65 to 200 mm (2.5 to 8 in.) (flange type, wafer type)
250 to 300 mm (10 to 12 in.) (flange type)
350 to 400 mm (14 to 16 in.) (flange type)
0
(32)
–10
(14)
F0201.ai
*1
PFA Lining, Integral Flowmeter (excluding high pres-
sure type (ASME Class 600 ange) and Hygienic Use)
Pressure
MPa (psi)
4 (580)
2 (290)
1 (145)
– 0.1 (–14.5)
– 40
(– 40)
40
(104)
100
(212)
Temperature °C (°F)
130
(266)
0
(32)
–10
(14)
2.5 to 50 mm (0.1 to 2 in.) (flange type, wafer type)
65 to 200 mm (2.5 to 8 in.) (flange type, wafer type)
250 to 300 mm (10 to 12 in.) (flange type)
350 to 400 mm (14 to 16 in.) (flange type)
F0202.ai
*1
*1: For wafer types of 150 to 200 mm (6 to 8 in.) or carbon
steel ange types (process connection code: C##) of 50 to
400 mm (2 to 16 in.), the minimum temperature is –10°C
(14°F).
PFA Lining, Remote Sensor and Integral Flowmeter
(high pressure type, ASME Class 600 ange)
F0203.ai
Pressure
MPa (psi)
Temperature °C (°F)
93.3
(200)
100
(212)
37.8
(100)
8.27 (1200)
8.15 (1182)
9.92 (1440)
50
(122)
0
(32)
-10
(14)
-0.1 (-14.5)
10 (1450)
PFA Lining, Hygienic Use (Remote Sensor and Integral
Flowmeter)
F0204.ai
Temperature °C (°F)
Pressure
MPa (psi)
1(145)
–0.1(–14.5)
15 to 125 mm (0.5 to 5 in.) (integral flowmeter)
150
(302)
130
(266)
15 to 125 mm (0.5 to 5 in.) (remote sensor)
0
(32)
–10
(14)
Ceramics Tube, Remote Sensor
Pressure
MPa (psi)
Temperature °C (°F)
2.5 to 50 mm (0.1 to 2 in.)
80 to 200 mm (3 to 8 in.)
4 (580)
2 (290)
–0.1 (–14.5)
–10
(14)
0
(32)
100
(212)
150
(302)
F0205.ai
Ceramics Tube, Integral Flowmeter
Pressure
MPa (psi)
4 (580)
3.25 (471)
2 (290)
–0.1(–14.5)
2.5 to 50 mm (0.1 to 2 in.)
80 to 200 mm (3 to 8 in.)
–10
(14)
0
(32)
100
(212)
130
(266)
Temperature °C (°F)
F0206.ai