Code Message Explanation/Countermeasures/Ref.section
96 This menu is locked. Check the key lock setting. Release the key lock.
See section 8.1.
97 This function is locked. To use the function, login as a user who has permission
to use it.
See section 8.2 in the advanced security manual.
100 IP address doesn’t belong to class A, B, or C. Check the IP address.
See section 1.3 in the communication manual.
101 The result of the masked IP address is all 0s or 1s. Check the subnet mask.
See section 1.3 in the communication manual.
102 SUBNET mask is incorrect. Check the subnet mask.
See section 1.3 in the communication manual.
103 The net part of default gateway is not equal to that of IP
address.
Check the IP address.
See section 1.3 in the communication manual.
105 This port number is already in use. Please enter a different
number.
Enter a different port number for each function.
See section 6.1 in the communication manual.
110 This user name is not registered. Log in as a registered user.
See section 8.2 and section 2.1 in the advanced
security manual.
111 The login user ID is incorrect. Enter the correct user ID.
See the advanced security manual.
112 Password must use more than 6 alphanumeric characters. The password must be between 6 and 20 characters in
length.
See section 2.1 in the advanced security manual.
113 Password entered is incorrect. Enter the correct password.
See section 8.3 and section 2.2 in the advanced
security manual.
114 This user name is invalid. Have a system administrator clear the “invalid user”
status.
See the advanced security manual.
116 This user name cannot be specified. The user name is reserved.
See section 8.2 and the advanced security manual.
117 This password is not effective. Because the wrong password has been entered
consecutively for more than the permissible number of
times, this user is invalid. Have an administrator clear
the “invalid user” status.
See section 2.2 in the advanced security manual.
119 This user name is unable to use this mode.
A user-level user cannot enter the Basic Setting mode
See section 8.2.
120 Measured value is incorrect. (in ascending order) Set the calibration correction value to a value greater
than the previous value.
See section 3.9.
121 A user is already logged in. Another user has logged in with the same name or is
logging in.
See section 1.3 in the advanced security manual.
122 Measured value exceeds the range setting. Check the channel input or the channel range setting.
See sections 3.3 and 3.9.
124 Password entry cannot be performed. You cannot use bar codes to enter a password.
See section 2.9 in the advanced security manual.
125 Character entry cannot be performed. The DX is not showing a display used to enter character
strings.
See section 2.11.
126 You cannot use the same password. Specify a different password.
See section 8.3.
127 Report kind overlaps and cannot be set up. Change the overlapped report data type.
See section 9.5.
128 “Logout” cannot be set to “Hide”. See sections 5.17 and 8.3.
129 IP address is not set. Set the IP address of the DX.
See section 1.3 in the communication manual
131 You have exceeded the available channel capacity. You cannot connect more than 240 channels.
See section 1.10 in the communication manual