11-16
IM 04L42B01-01E
• OtherCommunicationErrors
An English error message is returned via the communication interface. It is not displayed
on the screen.
Code Message
400 Input username.
Enter the name of a user that is registered on the DX and that is allowed to log in to the DX through communication.
401 Input password.
Enter the password.
402 Select username from ‘admin’ or ‘user’.
Select “admin” or “user” to specify the user level to connect to.
403 Login incorrect, try again!
The login failed. Enter the information again, starting with the user name.
404 No more login at the specified level is acceptable.
Connecting would cause the maximum number of simultaneous connections (administrator: 1, normal user: 2 or
setting: 1, monitoring: 2) to be exceeded. Connect to a different level, or exit by entering “quit.”
405 Input user ID.
Enter a user ID.
406 Select function from ‘setting’ or ‘monitor’.
Select “setting” (to connect to the setting function) or “monitor” (to connect to the monitoring function).
407 Password has expired. Please enter a new password.
The password has expired. Enter a new password.
408 Enter password again for confirmation.
Enter the password again for confirmation.
409 This password is not correct or was already used.
The password is incorrect or has already been used.
410 Login successful. (The special user level)
You have logged in to the administrator level.
411 Login successful. (The general user level)
You have logged in to the user level.
420 Connection has been lost.
The connection has been lost.
421 The number of simultaneous connection has been exceeded.
The maximum number of simultaneous connections has been reached, so no further connections can be made.
422 Communication has timed-out.
Communication timed out so the connection was closed.
450 This entry is incorrect.
You cannot currently log in. This may happen for one of the following reasons:
•Theusernameisnotregistered.
•Theusernameisinvalid.
•TheuserIDiswrong.
451 Login prohibited because another user is logged in.
Another user is already logged in. This may happen for one of the following reasons:
•Someonehasalreadyloggedinwiththesameusername.
•Whenthemulti-loginfunctionisnotenabledandsomeonehasloggedinthroughkeyoperations.
452 This command must be used with LL command.
Use the LL command.
11.1AListofMessages

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA Daqstation DX2000 and is the answer not in the manual?

YOKOGAWA Daqstation DX2000 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDaqstation DX2000
CategoryRecording Equipment
LanguageEnglish

Related product manuals