20
IM 04L21B01-02EN
5.
Press DISP/ENTER once.
Select First-CH.
6.
001 is displayed next to First-CH, so do not change this setting.
Press the +1 soft key once to set First-CH and Last-CH to 002.
7.
Press the down arrow key once to move the cursor to Mode.
8.
Press the TC soft key once. The cursor moves to Range, and the changed item is
displayed in yellow.
Select TC.
9.
Press the Next soft key.
10.
Press the T soft key once. The cursor moves to Span_L.
11.
Press the Input soft key once.
Display the Span_L setting window.
12.
Enter 0.0 in the Span Lower box.
Select the input position: Press the
right arrow key
once to move the cursor in the
text box to the right.
Delete the minus sign: Press the
up arrow key
once to delete the minus sign.
Delete the 2 and 0 in the same way.
Enter the input: Press
DISP/ENTER
once.
Span_L
is set, and the cursor
moves to
Span_U
.
Cancel the setting: Press
ESC
before pressing
DISP/ENTER
(entering the
input).
13.
Enter 200.0 in the Span Upper box.
See step 12 for the procedure.
Setting the Functions and Operations

Other manuals for YOKOGAWA FX1012

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA FX1012 and is the answer not in the manual?

YOKOGAWA FX1012 Specifications

General IconGeneral
BrandYOKOGAWA
ModelFX1012
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals