EasyManuals Logo

YOKOGAWA MVAdvanced MV1000 User Manual

YOKOGAWA MVAdvanced MV1000
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
3-6
SM MV1000-01E
Measurement Accuracy
● Overview
The measurement accuracy is tested using two different* integral times.
* But only one integral time is used when using the MV2000 with the /MC1 option.
The measurement accuracy test is performed in one of the following ways.
One representative channel for each A/D converter is tested, and the channel-to-
channel error between the representative channel and other channels is tested (see
next page).
* Choose any channel number for each A/D from those listed in the A/D converter
configuration table.
Ex.: With the MV1004, set channels 1 and 3 as representative channels.
The measuring accuracy of all channels is tested separately.
● A/DConverterConfiguration
Models
Measurement
Interval
Number of
Channels
No. of
A/Ds
Channel Number by A/D
MV1004 125 ms 4CH 2 1-2, 3-4
MV1006 1 s 6CH 1 1-6
MV1008 125 ms 8CH 4 1-2, 3-4, 5-6, 7-8
MV1012 1 s 12CH 1 1-12
MV1024 1 s 24CH 2 1-12, 13-24
MV2008 125 ms 8CH 4 1-2, 3-4, 5-6, 7-8
MV2010 1 s 10CH 1 1-10
MV2020 1 s 20CH 2 1-10, 11-20
MV2030 1 s 30CH 3 1-10, 11-20, 21-30
MV2040 1 s 40CH 4 1-10, 11-20, 21-30, 31-40
MV2048 1 s 48CH 4 1-12, 13-24, 25-36, 37-48
● TestInstruments
Instrument Specifications
DC voltage generator Accuracy: ±0.005% of setting + 1 µV
DMM Accuracy: ±0.005% of rdg + 1 µV
Variable resistors Accuracy: 0.01% + 2 mΩ or better
● Setup
Set the input range of the channels under test, and the test range.
● TestProcedures
Warm up the instrument for thirty minutes before performing the test.
Enter the input value determined for each range. Check whether the displayed
measured values satisfy the reference values.
DC voltage generator
To the input
terminals
+
Variable resistors
The resistance of three
lead wires must be equal.
Input terminal block
3.4 Test Procedures

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MVAdvanced MV1000 and is the answer not in the manual?

YOKOGAWA MVAdvanced MV1000 Specifications

General IconGeneral
BrandYOKOGAWA
ModelMVAdvanced MV1000
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals