5. Remove the display from housing by slowly pulling forward.
ð The switch can be seen.
6. Set the desired address.
OFF
1 2
H
L
SW1
ON
0
3
1
0
127
6
5
4
Address
1 2
ON
OFF
1 2
SW2SW3
7. Push display into housing.
8. Fasten the display with 2 screws.
9. Screw display cover back onto transmitter housing.
10. Tighten the locking screw by turning it counterclockwise with an Allen wrench (size:
3.0).