< 2. Speci cations >
2-1
IIM 12A0V2-E
2. SPECIFICATIONS
Measurement:
Free available chlorine concentration of tap water, pH of or tap water industrial water.
Instruments integrated:
FC400G non-reagent type free available chlorine analyzer, PH400G or PH450G 4-wire pH converter
system, Refer to respective user’s manual.
Calibration method:
Chlorine analyzer: Manual/auto zero calibration
pH meter: Manual calibration
Cleaning system:
Chlorine analyzer: Not provided
pH analyzer: Ultrasonic cleaning
Sample (process) conditions:
Pressure: 100 to 750 kPa (Under 100 kPa, consult Yokogawa)
Flow rate: Chlorine analyzer: 0.1 to 10 l/min
pH analyzer: 3 to 10 l/min
Temperature: 0 to 50 °C
Ambient conditions:
Temperature: -5 to 50 °C
(Protection against freeze should be provided when necessary in winter.)
Humidity: 5 to 95% RH, non-condensing (10 to 90 % for pH converter system)
Storage temperature:
-30 to 70 °C (-25 to 70 °C when using PUS400G)
Dimensions:
Chlorine analyzer integrated: 530(W) × approx. 1312(H) × 550(D) mm
pH analyzer integrated: 530(W) × approx. 1320(H)*× 550(D) mm
Both integrated: 530(W) × approx. 1320(H)*× 550(D) mm
*) Approx. 1370(H) when option code of “/TT3” speci ed.
Materials:
Piping: Hard PVC, polypropylene resin
Stanchion: Carbon steel or stainless steel
Finish: Baked polyurethane resin coating
Color: Equivalent to Munsell 0.6GY3.1/2.0