How to resolve Err. 7 (Response Error during Zero and / or Span Calibration) on YOKOGAWA Vigilantplant ST401G Measuring Instruments?
- ZZachary OliverAug 20, 2025
To resolve Err. 7, confirm response recovery after polishing the indicator electrode with alumina. Also, check that the stability determination parameters “*?T. SEC”, “ *?PPM ”, and “*CT.min” set in service code 04 are correct. If the parameters are correct but there is still no response recovery, the indicator electrode may have deteriorated and should be replaced.