C
Christopher RichardsonSep 2, 2025
How to fix Err. 2 (Sample Temperature Error) on YOKOGAWA Vigilantplant ST401G Measuring Instruments?
- CcastrochristopherSep 2, 2025
To fix Err. 2 (Sample Temperature Error), control the sample temperature to be in the range of 0.0 to 50.0°C. If the error persists within this range, the RTD may be defective (open circuited), and the counter electrode should be replaced.