IM 253401-01E
App2-44
Appendix 2.7 Communication-related Error Messages
Errors in communications execution (200 to 299)
Code Message Action Reference Page
221 Setting conflict Check the relevant setting. App.2.3
222 Data out of range Check the setting range. App.2.3
223 Too much data Check the data byte length. App.2.3
224 Illegal parameter value Check the setting range. App.2.3
241 Hardware missing Check availability of options. —
260 Expression error Equation is not allowed. —
270 Macro error Does not conform to the macro definition specified in IEEE488.2. —
272 Macro execution error Does not conform to the macro definition specified in IEEE488.2. —
273 Illegal macro label Does not conform to the macro definition specified in IEEE488.2. —
275 Macro definition too long Does not conform to the macro definition specified in IEEE488.2. —
276 Macro recursion error Does not conform to the macro function specified in IEEE488.2. —
277 Macro redefinition not allowed Does not conform to the macro definition specified in IEEE488.2. —
278 Macro header not found Does not conform to the macro definition specified in IEEE488.2. —
Error in communication Query(400 to 499)
Code Message Action Reference Page
410 Query INTERRUPTED Check transmission/reception order. App2-3
420 Query UNTERMINATED Check transmission/reception order. App2-3
430 Query DEADLOCKED Limit the length of the program message including <PMT> to App2-4
1024 bytes or less.
440 Query UNTERMINATED after Do not enter any query after *IDN? and *OPT?. —
indefinite response
Errors in Execution (800 to 899)
Codes Message Action Reference Page
813 to 819 Invalid operation Refer to 14.4 14-11
830 to 833 Internal memory access error Refer to 14.4 14-11
841 to 847 Integrator execute error Refer to 14.4 14-11
Error in System Operation (912)
Code Message Action Reference Page
912
Fatal error in Communication driver
Service is required. —
Warnings (350, 390)
Code Message Action Reference Page
350 Queue overflow Read out the queue. App2-39
390 Overrun error Adjust the baud rate. —
(only for RS-232C)
Note
The warning code 350 only appears in case of an overflow of the error queue. The error which occurs in case
of clearing the
STATus:ERRor?
will not appear on the screen.