Tristel Duo NCU
UK
Chlorine dioxide
Tristel trio wipes system
UK
Enzymes, Chlorine Dioxide
Accel® PREVention™ Wipes & RTU
CA
Hydrogen Peroxide
CaviWipes
US
isopropyl alcohol
Sani-Cloth HB Germicidal Disposable
Wipe
US
Alcohol free quaternary ammonium
Sani-Cloth Plus Germicidal Disposable
Cloth
US
isopropyl alcohol
Cautio
n
1
.
The probe must be cleaned after each use;
2
.
Do not use surgical brush to clean the probe. Even soft brush may
damage the probe. Only soft cloth is allowed;
WARNING
1
.
The product must be away from machine with high current or
strong magnetic.
2
Do not immerse the probe in any type of liquid or detergent.
3
.
Disinfection of probe with gas or heating is strictly
prohibited.
4
.
It is strictly prohibited to immerse probe socket in any
solution. Probe surface is waterproof but only probe head.
Please be careful while using coupling agent and cleaning
or disinfecting probe surface.
5
.
Users are not allowed to repair without authorization in
case of probe failure, Product must be sent back to the
company for repair.
Check the following items after cleaning and disinfection
Inspection
items
Inspection methods
Acceptance Criteria
Acoustical lens
Visual inspection of lens surface and connection
between lens and shell of the acoustic head
No damages or cracks on lens surface;
No cracks at connection between lens
and shell.
Shell
Visual inspection of shell surface
No damages or cracks on shell surface
Shell
connection
Visual inspection of shell connection
No cracks or lack of gluing at joints of
the shell
Residue
Visual inspection of probe surfaces, especially
pits and narrow areas
No residual detergent or disinfectant