13
5. Under Step 2 of the Text Import Wizard, choose Semicolon.
6. Under Step 3, click on the column with the incorrectly formatted
data. This column should be highlighted in black. Choose Text
under Column data format. Do this for each column with incorrectly
formatted data.
7. Select Finish, then choose where you want the data to be placed on
your opened spreadsheet.
MAINTENANCE
Dissolved Oxygen Sensor
Clean the probe whenever it won’t calibrate or the readings are unstable
and changing the membrane does not solve either issue. On average, the
probe should only be cleaned about once or twice per year. To clean the
probe, sand the electrodes according the directions below. In addition to
606035REF-B-YSI-DO200A-DO200M-User-Manual-English.indd 13 12/7/2017 9:45:12 AM