To begin:
Insert the Test Eye (Figure 1) in the slots on the chin rest (Figure2).
Log into the ATLAS software as the Administrator.
Leave the password box blank.
Select Tools from the menu and then select Calibration Check and align the projection head by
centering the crosshairs in the smallest ring of the image displayed (Figure 3).
How can I perform a calibration test on my ATLAS 9000?
ZEISS Quick Help: ATLAS
®
9000
A calibration test can be performed on your ATLAS 9000 using a calibration tool that is often
referred to as the Test Eye. Below is the 3-step check to perform a calibration test.
Figure 1 Figure 2
Figure 3