The same refers to pseudophakic eyes if the intraocular lens does not
sufficiently scatter the light from the slit.
Fig. 65Ą Optical section of a pseudophakic eye. On the right, the unsharp reflection
of the rear face of the IOL is visible.
Under certain circumstances, a strong reflection may be visible on pseuĆ
dophakic eyes, which is produced by the rear surface of the lens.
When you release an anteriorĆchamber depth measurement in such a
case, the evaluation software will interpret the bright reflection as the
front surface of the crystalline lens. The instrument will then display a
measured value in the range of about 5 to 6 mm. This value is not the
exact anterior chamber depth!