14.1 Safety
CryoMini is manufactured according to the safety regulations DIN EN 60601-1.
As manufacturer, Zimmer MedizinSysteme can only be considered responsible for the
safety and reliability:
• the device is operated using a proper power outlet with earth contact and the
electrical installation complies with DIN VDE 0100 Part 710,
• the equipment is operated in accordance with the instructions for use,
• extensions, readjustments or modifications are carried out only by persons
authorised by Zimmer MedizinSysteme GmbH,
• the user has ascertained the functional safety, the proper operating condition and
mechanical integrity of CryoMini before use,
• the device is operated only by properly trained personnel,
• the device is not operated in hazardous areas and / or a combustive atmosphere,
• the device is immediately disconnected from the mains when penetrated by liquid.
The device does not contain any parts that must be maintained or repaired by the
operator.