Manufacturer´s EMC declaration
Guidelines and manufacturer’s declaration – electromagnetic immunity
The CryoMini device is intended for use in the electromagnetic environment specified below. The customer or the user of the
CryoMini device should assure that it is used in such an environment.
Electromagnetic immunity
tests
Electromagnetic environment -
guidelines
(ESD) acc. to IEC 61000-4-2
± 8 kV air discharge
± 8 kV air discharge
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative air humidity must be
Electrical fast transient
disturbance variables /
bursts acc. to IEC 61000-4-4
lines
± 1 kV for input / output
lines
Not applicable for input
Mains power quality should be
that of a typical commercial or
hospital environment.
according to IEC 6100-4-5
voltage
± 2 kV common-mode
voltage
± 2 kV common-mode
Mains power quality should be
that of a typical commercial or
hospital environment.
interruptions and voltage
variations of power supply
acc. to IEC 61000-4-11
T
(>95% dip in UT for ½
cycle)
40% U
T
(60% dip in U
T
for 5 cycles)
70% U
T
(30% dip in U
T
for 25
cycles)
<5% U
T
(>95% dip in UT for 5
seconds)
T
(>95% dip in UT for ½
cycle)
40% U
T
(60% dip in U
T
for 5
cycles)
70% U
T
(30% dip in U
T
for 25
cycles)
<5% U
T
(>95% dip in UT for 5
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of CryoMini requires continued
operation also during power
mains interruptions, it is
recommended that CryoMini be
powered from an uninterruptable
power supply or a battery.
Magnetic field at the power
frequency (50/60 Hz)
according to IEC 61000-4-8
Magnetic fields at this power
frequency should be at levels
characteristic of a location in a
typical commercial or hospital
Note: U
T
is the AC mains voltage prior to application of the test level.