21
Warnings:
1. Place Safe/Run Slide in the "safe" position when not in use.
2. Always operate the Low Speed Drill with the appropriate
attachment.
3. Check each bur before use in the Low Speed Drill. Never use a
bent bur.
Note: The Low Speed Drill accepts standard dental latch burs or
Hall
®
Surgical burs of the 5088 series.
The Low Speed Drill operates with two different head attachments. When
the Drill Attachment is in use, the handpiece will operate at a maximum
speed of 2,500 rpm with 15 in. oz. of torque. The Screw/Tap Attachment
operates with a maximum speed of 30 rpm. In the Screw function there are
25 in. oz. of torque, while in the Tap mode there are 55 in. oz. of torque in
the forward direction and 60 in. oz. of torque in the reverse direction.
Note: Low Speed Drill torques while in use:
Drill mode 15 in. oz. will be maintained.
Screw mode If 25 in. oz. is exceeded the
handpiece will stop and
MAXIMUM TORQUE
will be
displayed. The handpiece can
only be reset by secondary
activation of the footswitch.
Tap mode 55/60 in. oz. will be maintained.
Note: The Low Speed Drill can only be operated by the Footswitch.
Warning: Place handpiece in the "safe" position prior to connection
with the attachment. Be certain the attachment/accessory is completely
seated on the handpiece prior to each use.