LEVERLESS OPERATION
The MicroChoice™ High Speed Drill, Medium Speed Drill, Reciprocating
Saw, Sagittal Saw and Oscillating Saw can also be operated with the
Footswitch.
To remove the Handpiece Activation Lever:
1. Place the Safe Slide in the "safe" position.
2. Depress the Activation Lever
on the ribbed portion.
3. At the same time, push the
Activation Lever towards the
Safety Slide until it disengages
from the handpiece. For a brief
period, as the Activation Lever
is disengaged from the
handpiece, the MicroChoice
Controller will display
MAGNETIC FIELD.
4. To reattach the Activation
Lever, reverse the steps above.