55
4. To dry, wipe surfaces with a
clean, lint-free towel.
5. It is best to use burs and blades
only once. However, if burs or
blades are reused, clean
cutting surfaces with a wire
brush and mild detergent. Be
certain that all surfaces are
free of debris. Rinse with
running water. Check burs and
blades for dull or chipped
surfaces.
6. To clean the cannulated section of Wiredrivers:
* Remove the wire guard.
* Feed the wire end of the
cleaning brush through
the back of the handpiece.
* Pull the brush through the
handpiece. Repeat until
debris is removed.
7. For care, cleaning, lubrication and maintenance instructions of bur
guards and attachments/accessories, check the package instruction sheet
that accompanies the attachment/accessory. If the instruction sheet
cannot be located, contact Hall
®
Surgical for information concerning
these issues. Replacement instruction sheet(s) are available.