EasyManua.ls Logo

Zodiac TRi pH Module - Calibrating the Sensor(S); Adjusting the Set Point(S)

Zodiac TRi pH Module
14 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
H0412500.A - EN - 2013-02
8
EN
- TRi PRO module:
pH set point
ACL set point
A '^' symbol appears to the right of the set point if the value measured by the equipment needs to be corrected automa-
cally (pH of water in the pool higher than the set point and/or insucient chlorine level). Then pH minus is injected and/
or chlorine produced automacally according to the dened cycles.
By default, pH regulaon (peristalc pump in the TRi pH or TRi PRO module) is disabled and the LCD screen will
display 'pH ---'. It is enabled automacally about 8 hours aer being turned on. To enable pH regulaon immedi-
ately and so display the set point on the LCD screen, see § 3.6.2.
3.3 Calibrang the sensor(s)
To operate accurately and reliably, the sensors must be calibrated before using the TRi chlorinator with the TRi
pH or TRi PRO module. To maintain maximum eciency of the unit, we recommend calibrang at least once
every 2 months during the period the swimming pool is being used.
Ensure the sensors are cleaned before calibraon (see § 4.1).
Conrm that the TRi power pack is plugged in,
Stop the pool pump and isolate the sensors by closing any valves, so as to be able to re-
move the sensors safely.
Remove the sensors from their sensor holder on the POD (unscrew the locking ring by
hand).
Rinse the end of the sensors with clean water and shake them to remove excess water.
Do not touch or wipe the glass bulb at the end of the sensors.
Put the pH sensor into a sample of the pH7.5 buer soluon supplied.
Put the ACL sensor into a sample of the 700 mV buer soluon supplied.
Leave the sensors immersed for about 1 minute to obtain a reliable measurement.
Press , then use the or keys to display 'CALIB. PH' or 'CALIB. PH/ACL', then press .
Use the or keys to display 'PH CALIBRATION' or 'ACL CALIBRATION', then press .
Note the measured value:
pH = 'X.X'
- If the value is greater than 8.2 or less than 6.4: the sensor is dirty or has been damaged during transport. Clean the pH
sensor (see § 4.1) then repeat the above steps. If the problem persists, contact your reseller.
- If the value is between 6.4 and 8.2, press to begin calibraon. The procedure takes about 15 seconds.
- When calibraon is nished, check that the value displayed equals 7.5, otherwise repeat the calibraon.
ACL = 'XXX'
- Start the calibraon procedure by pressing ; the procedure lasts about 15 seconds,
- If the value is 700: calibraon is correct
- If the value is not 700: repeat the calibraon steps. If the problem persists, contact your reseller.
Press or wait 30 seconds to exit.
3.4 Adjusng the set point(s)
The set point is displayed connuously on the LCD home screen.
The default seng for the pH set point is pH7.2 (TRi pH and TRi PRO modules).
The default seng for the ACL set point is 4 (TRi PRO module).
This value provides the ideal compromise to achieve opmum eciency in terms of water disinfecon.

Related product manuals