9
7. Specifications
7.1. 3M™Steri-Vac™ Sterilizer/Aerator GS Series StructuralSpecifications
GS5 GS8
Shipping Weight
163kg (359lbs.) singledoor
169kg (373lbs.) double door
355kg (782lbs.) singledoor
362kg (799lbs.) doubledoor
Oper
ational Weight
127kg (281lbs.) singledoor
132kg (290lbs.) double door
261kg (576lbs.) singledoor
269kg (593lbs.) doubledoor
Exterior Dimensions
H 70.9cm x W 76.2cm x D 95.0cm
H 27.9in. x W 30.0in. x D 37.4in.
H 179.8cm x W 94.0cm x D 109.0cm
H 70.8in. x W 37.0in. x D 42.9in.
Chamber Internal Volume 136L (4.8cubic feet) 224L (7.9cubicfeet)
Chamber Internal Dimensions
H 38.0cm x W 43.0cm x D 83.0cm
H 15.0in. x W 17.0in. x D 32.5in.
H 46.0cm x W 51.0cm x D 97.0cm
H 18.0in. x W 20.0in. x D 38.0in.
7.2. Sound Levels
The average decibel level of an active GS5 or GS8 sterilizer is <60 dBA. For more information on sound levels, in the US contact 3M Health Care Helpline at 1-800-228-3957. Outside the US
contact your local 3M office or to locate your local office go to www.3M.com
7.3. PowerSpecifications
WARNING: To reduce the risk of shock due to hazardousvoltage,
the customer must provide a properly grounded outlet (an earth ground) for installation as described in the installation requirements section of this manual. Do not use a detachable cord rated
less than 15A.
CAUTION: To reduce patient risks associated with exposure to potentially non-sterile devices or inadequate sterilizationprocedures,
do not place any device emitting strong electronic magnetic fields (EMFs) near thesterilizer.
Electrical Power Operating Condition Units
Voltage Range 200- 240 VAC
Frequency 50/60 Hertz
Phase Single NotApplicable
GS5Current 10 Amps
GS8Current 12 Amps