Questions and Answers

  • D
    Daniel KingAug 17, 2025
    What to do if there is an L4 error after EO gas is in the chamber for 3M Steri-Vac GSX Laboratory Equipment?
    • S
      Scott JohnsonAug 17, 2025
      If an L4 error occurs after EO gas is in the chamber, a Supervisor or Service personnel must clear the error. Wait until the Stop button is red. The Supervisor or Service personnel should press Stop Cycle and enter the PIN, then press Clear and enter the PIN to clear the error. Remember to document everything per facility procedures.
  • S
    Scott WardAug 19, 2025
    How to clear L1 error before EO gas in chamber for 3M Steri-Vac GSX?
    • J
      jbeardAug 19, 2025
      To clear an L1 error that occurs before EO gas is in the chamber, the operator can clear the error and access the load. First, wait until the Open Door or Open Load Door button is blue. Then, press Open Door and press Clear to clear the error. Make sure to document everything per facility procedures and restart the cycle per facility procedures.
  • K
    keithsantosAug 22, 2025
    What to do if there is a C604 power interruption in 3M Steri-Vac GSX Laboratory Equipment?
    • M
      Michelle JohnsonAug 22, 2025
      In the event of a C604 power interruption, document the issue per facility procedures.
  • J
    James MitchellAug 25, 2025
    What to do if there is an E702 user abort error on 3M Steri-Vac GSX Laboratory Equipment?
    • T
      Timothy PriceAug 26, 2025
      If an E702 user abort error occurs, clear the details.
  • J
    Jason HallAug 28, 2025
    What to do if 3M Steri-Vac GSX shows an L2 error before EO gas, requiring service?
    • M
      Mrs. Heather Chavez PhDAug 28, 2025
      If an L2 error occurs before EO gas is in the chamber and requires service, the operator can access the load. Wait until the Open Door or Open Load Door button is blue, then press Open Door. Call Service, as Service must clear the error. Be sure to document the issue per facility procedures and restart the cycle per those procedures.
  • D
    Daniel BautistaAug 30, 2025
    What to do if 3M Steri-Vac GSX has an L5 error after EO gas requiring service?
    • J
      Jason RobinsonAug 30, 2025
      If an L5 error occurs after EO gas is in the chamber and requires service, call Service. Service must access the load and clear the error. Remember to document everything per facility procedures.
  • J
    Jessica McclainSep 2, 2025
    What should I do if there's an L6 error after EO gas in the chamber, requiring service attention for 3M Steri-Vac GSX Laboratory Equipment?
    • C
      Charles WangSep 2, 2025
      If an L6 error occurs after EO gas is in the chamber and requires service attention, call Service. Service must access the load and clear the error. Document the issue per facility procedures.
  • B
    brandysmithSep 5, 2025
    What to do when 3M Laboratory Equipment shows A2 errors in temperature control during programmed aeration?
    • S
      Sabrina MurphySep 5, 2025
      If A2 errors occur in temperature control during programmed aeration, call Service. Service must access the load and clear the error. Document per facility procedures.
  • T
    Troy OrozcoSep 8, 2025
    What to do if there is an L3 error before EO gas requiring service for 3M Steri-Vac GSX?
    • R
      randybenitezSep 8, 2025
      If an L3 error occurs before EO gas is in the chamber, requiring service before the operator can open the load door, call Service. Service must access the load and clear the error. Document the issue and restart the cycle per facility procedures.
  • C
    Carrie PenaSep 12, 2025
    What to do when 3M Steri-Vac GSX Laboratory Equipment has L7 errors requiring 3M Service?
    • M
      Megan LunaSep 12, 2025
      If L7 errors occur requiring 3M Service, call Service. Service must clear the error. Document per facility procedures.