12.2.2. Non‑compatiblePackaging
The following packaging is not compatible and is not recommended for use in ethylene oxide (EO) sterilizationprocesses:
Do not use packages made entirely of any of the followingmaterials:
• Foil
• Cellophane
• Polyvinylchloride(PVC)
• Impervious polypropylenefilm
• Polyester (polyester film made from stretched polyethylene terephthalate (PET))
• Polyamide(nylon)
• Polyvinylidenechloride
12.2.3. Package MedicalDevices
1�  Place the device in ethylene oxide (EO) compatiblepackaging.
2�  If applicable per your facility’s procedures, place one or more EO chemical indicators (CI) inside each container, wrapped tray, or sterilization pouch (Chapter14: Process Monitoring, Load 
Release, and SterilizerQualification).
3�  When sterilizing pouched items, seal the sterilization pouch using a heat sealer set to the appropriate time andtemperature.
4�  Wrap non‑containerized trays with EO compatible wrapping material and secure with EO indicatortape.
5�  For containerized items, ensure an external chemical indicator for EO ispresent.
NOTE: When loading devices inside a container or tray, leave sufficient space between the instruments to enable the sterilant to circulate. The containers used must be prepared and sealed 
according to the procedures recommended by their respectivemanufacturers.
12.3. Loading the 3M™Steri‑Vac™ Sterilizer/Aerator GSXSeries
Loading baskets are recommended to facilitate loading the sterilization chamber. Loading baskets are designed with skids to avoid scratching the chamber floor. A transport cart can be used to 
easily load, unload, and move loading baskets to and from the 3M™Steri‑Vac™ Sterilizer/Aerator GSXSeries.
 DANGER: To reduce the risks associated with exposure to ethylene oxide,
do not overload the sterilization chamber. Use good practices for loading the 3M™Steri‑Vac™ Sterilizer/Aerator GSX Series chamber.
 CAUTION: To reduce the risk of injury,
follow good ergonomic practices. Loading baskets should not be overfilled requiring excessive force in pulling and pushing loaded baskets in and out of the 3M™Steri‑Vac™ Sterilizer/
Aerator GSX Series chamber. Reference facility policies and procedures for appropriate ergonomic practices. 
 CAUTION: To reduce patient risks associated with exposure to potentially non‑sterile devices or inadequate sterilization procedures:
Always use chemical indicators and biological indicators for monitoring the performance of sterilization cycles as described in this manual. Always use chemical indicators and biological 
indicators per the device manufacturer’s instructions for use (IFU).
Do not overload the sterilization chamber. Use good practices for loading the 3M™Steri‑Vac™ Sterilizer/Aerator GSX Series chamber.