11.11. CycleReports
11.11.1.Custom Cycle SetPoints
The 3M™Steri‑Vac™ Sterilizer/Aerator GSX Series allows users to view and print parameter values for custom cycles installed on the GSX Series sterilizer. The report will display programmed
custom cycle set‑points for eachparameter.
1 To view or print custom cycle set points go to Menu>Reports>Custom cycle setpoints.
2 Select the desired cycle and then View or Print. Figure62is an example custom cycle set pointreport.
3 Press Cancel to return to the mainscreen.
Custom cycle report
EA343434 04/20/2015 15:13:27
Software version 1.0.15
--------------------------------------
Cycle: 1 Test cycle
Stage: Preheat
Cycle temperature = 60 °C(140 °F)
Timeout = 60 min
Hold time = 0 min
Stage: Air removal
Heatsink temperature = 95 °C(203 °F)
Chamber pressure = 160 mbar
Duty cycle = 100 %
Stage: Chamber test
Chamber test time = 5 min
Stage: Conditioning
Relative humidity = 40 %
Humidify timeout = 50 min
Hold time = 0 min
RH injection rate = Fast
RH threshold lower = 10 %
RH threshold upper = 90 %
Stage EO injection
Simulate EO = Yes
Inject pressure = 160 mbar
Minimum pressure increase = 200 mbar
RH threshold lower = 20 %
RH threshold upper = 60 %
Stage: EO exposure
Exposure time = 1 min
Stage: EO removal
Stage: Flushing
Number of purge cycles = 5
Vent duty cycle = 100 %
Dwell time top = 4 min
Pump down pressure = 300 mbar
Dwell time bottom = 0 min
Purge aeration time = 90 min
Mandatory locked aeration = 180 min
Stage: Aeration
Aeration temperature = 34 °C(93 °F)
Minimum aeration time = 0.5 hr
Figure62.
Example Custom Cycle Set PointReport