CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER  |  CI/FCB400/FCH400-EN REV. H  31 
 
 
A For vertical installation in a riser, no special measures are 
required. 
B For vertical installation in a downpipe, a piping constriction 
or an orifice must be installed below the sensor. Doing this 
prevents the sensor from draining during the measurement. 
 
Horizontal installation 
 
Figure 7:  Horizontal installation 
 
A For liquid measuring media and horizontal installation, the 
transmitter and terminal box must point upward. If a self-
draining installation is required, the sensor must be mounted 
at an incline of ≥ 30°. 
B Installing the sensor at the highest point of the piping leads 
to an increased number of measuring errors due to the 
accumulation of air or the formation of gas bubbles in the 
meter tube. 
 
 
 
 
 
 
Gaseous measuring media 
 
Observe the following points to avoid measuring errors: 
•  Gases must be dry and free of liquids and condensates. 
•  Avoid the accumulation of liquids and the formation of 
condensate in the meter tube. 
•  During operation, there must be no phase transitions in 
the measuring medium. 
 
If there is a risk of condensate formation when using gaseous 
measuring media, note the following: 
Ensure that condensates cannot accumulate in front of the 
sensor. 
If this cannot be avoided, we recommend that the sensor is 
installed vertically with a downward flow direction. 
 
Vertical installation 
For vertical installation, no special measures are required. 
 
Horizontal installation 
 
Figure 8:  Horizontal installation 
 
A For gaseous measuring media and horizontal installation, the 
transmitter and terminal box must point downward. 
B Installing the sensor at the lowest point of the piping leads 
to an increased number of measuring errors due to the 
accumulation of liquid or the formation of condensates in 
the meter tube.