48  CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER  |  CI/FCB400/FCH400-EN REV. H   
 
 
 
… 6  Electrical connections 
 
Electrical data for inputs and outputs 
 
 
Note 
When using the device in potentially explosive atmospheres, 
note the additional temperature data in Use in potentially 
explosive atmospheres on page 6! 
 
 
Power supply L / N, 1+ / 2− 
 
AC voltage 
Terminals  L / N
Operating voltage  100 to 240 V AC, 50 / 60 Hz
Power consumption  < 20 VA
 
DC voltage 
Terminals  1+ / 2−
Operating voltage  19 to 30 V DC
Power consumption  < 20 W
 
 
Requirements for inputs and outputs 
For reasons of electromagnetic compatibility (EMC), shielded 
cables should be used in certain output configurations; this is 
presented in the table below. 
The cable shielding must be inserted in the device, see 
Connection to integral mount design on page 56 and 
Connection to remote mount design on page 58. 
 
 
Use of shielded cables 
Transmitter  Plug-in card used  Terminal 
Uco/31/32 
41/42 
51/52 
V1/V2 
V3/V4 
Dual-
compartment 
housing 
Active digital 
output V1/V2 
— — —  X  — 
Active digital 
output V3/V4 
— — —  —  X 
Modbus V1/V2  —  —  —  X  — 
Profibus DP V1/V2  —  —  —  X  — 
Single-
compartment 
housing 
Active digital 
output V1/V2 
X X X  —  X 
Active digital 
output V3/V4 
X X X  X  — 
Modbus V1/V2*  —  —  —  X  — 
Profibus DP V1/V2*  —  —  —  X  — 
Ethernet V1/V2  X  X  X  X  X 
Ethernet V1/V2, 
POE V3/V4 
X X X  X  X 
X  Use shielded cables 
*  Use in remote mount design only 
 
 
 
 
 
 
Current output 32 / Uco, 31 / 32 (basic device) 
Can be configured for outputting mass flow, volume flow, 
density and temperature via on-site software. 
 
 
A  Current output 31 / Uco, active  B  Current output 31 / 32, passive 
Figure 29:  (I = internal, E = external, R
B
 = load) 
 
 
Permissible source voltage U
q
 for passive outputs in relation to load 
resistance R
B
 where I
max
 = 22 mA.  = Permissible range 
Figure 30:  Source voltage for passive outputs 
 
Current output  Active Passive
Terminals  Uco / 32 31 / 32
Output signal  4 to 20 mA or 4 to 12 to 20 mA switchable
Load R
B
 250 Ω ≤ R
B
 ≤ 300 Ω 250 Ω ≤ R
B
 ≤ 600 Ω
Source voltage U
* 
13 V ≤ U
 ≤ 30 V
Measuring error  < 0.1 % of measured value 
Resolution  0.4 µA per digit 
*  The source voltage U
q
 is dependent of the load R
B
 and must be placed in an 
additional area. 
 
For information on communication via the HART protocol, refer 
to HART® Communication on page 61.