40  CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER  |  CI/FCB400/FCH400-EN REV. H   
 
 
 
… 5  Installation 
 
Installing the plug-in cards 
 
 
 
 WARNING 
Loss of Ex Approval! 
Loss of Ex Approval due to retrofitting of plug-in cards on devices for use in potentially explosive atmospheres. 
•  Devices for use in potentially explosive atmospheres may not be retrofitted with plug-in cards. 
•  If devices are to be used in potentially explosive atmospheres, the required plug-in cards must be specified when the order is 
placed. 
 
 
Optional plug-in cards 
The transmitter has two slots (OC1, OC2) into which plug-in cards can be inserted to extend inputs and outputs. The slots are located 
on the transmitter motherboard and can be accessed after removing the front housing cover. 
 
 
Plug-in cards    Pos.  Description  Quantity*
  1  Current output, 4 to 20 mA passive (red) 
Order number: 3KQZ400035U0100 
2
  2  Passive digital output (green) 
Order no.: 3KQZ400030U0100 
1**
  3  Passive digital input (yellow) 
Order no.: 3KQZ400032U0100 
2
  4  Loop power supply 24 V DC (blue) 
Order no.: 3KQZ400031U0100 
1
  5  Modbus RTU RS485 (white) 
Order no.: 3KQZ400028U0100 
1
  6  Profibus DP (white) 
Order no.: 3KQZ400027U0100 
1
7  Ethernet 
Order no.: 3KQZ400037U0100 
1
  8  Power over Ethernet (POE) 
Order no.: 3KQZ400039U0100 
1
  9  Active digital output (white) 
Order no.: 3KQZ400056U0100 
1**
*  The ‘Number’ column indicates the maximum number of plug-in cards of the same type that can be used. 
*  Only one plug-in card of the active digital output type or passive digital output type can be inserted in Pos. 2.