12 CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
… 2 Use in potentially explosive atmospheres
Temperature data
Temperature resistance for the connecting cable
The temperature at the cable entries of the device depends on
the design, the measuring medium temperature T
medium
and the
ambient temperature T
amb.
.
For the electric connection of the device, use only cables with
sufficient temperature resistance in accordance with the
following table.
Devices in integral mount design with dual-compartment
housing
T
amb.
Temperature resistance
50 °C ( 122 °F) 70 °C ( 158 °F)
60 °C ( 140 °F) 80 °C ( 176 °F)
70 °C ( 158 °F) 90 °C ( 194 °F)
Devices in integral mount design with single-compartment
housing
T
amb.
Temperature resistance
50 °C ( 122 °F) 75 °C ( 167 °F)
60 °C ( 140 °F) 85 °C ( 185 °F)
70 °C ( 158 °F) 95 °C ( 203 °F)
Sensor in remote mount design
T
amb.
Temperature resistance
50 °C ( 122 °F) 105 °C ( 221 °F)
60 °C ( 140 °F) 110 °C ( 230 °F)
70 °C ( 158 °F) 120 °C ( 248 °F)
For sensors in remote mount design, the wires in the connection
box must be additionally insulated with the enclosed silicone
hoses starting from ambient temperatures of T
amb.
60 °C
( 140 °F).
Environmental and process conditions for model
FCx4xx…
Ambient temperature T
amb.
-20 to 70 °C
(-4 to 158 °F)
-40 to 70 °C*
(-40 to 158 °F)*
Measuring medium temperature
T
medium
-40 to 205 °C
(-40 to 400 °F)
IP rating / NEMA rating IP 65, IP 67 /
NEMA 4X,Type 4X
* Optional, with order code ‘Ambient temperature range – TA9
Change from two to one column