EasyManua.ls Logo

ABB EL3060 Series - Data Transfer and Software Update Execution

ABB EL3060 Series
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 105
5 Click the command "Receive Data" in the "File" menu or click the icon
in the toolbar.
The configuration data is transferred from the gas analyzer to the com-
puter.
6 It is recommended that a back-up copy of the configuration data is
made before carrying out a software update:
Click the command "Save" or "Save as…" in the "File" menu
in the
toolbar and store the configuration data under a suitable file name.
7 Click the command "Software Update…" in the "Options" menu or click
the icon
in the toolbar.

Table of Contents

Related product manuals