EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 137
Troubleshooting
Flow error
Gas lines or filter fouled, clogged up or leaky
Disconnect gas analyzer from the gas conditioning system.
Blow compressed air through gas lines or mechanically clear them.
Exchange filter elements and filter beds.
Check gas lines for seal integrity.
Gas feed paths in the gas analyzer kinked or leaky
Disconnect gas analyzer from the gas conditioning system.
Check whether the gas lines in the gas analyzer are kinked or have be-
come detached from the connections.
Check gas feed paths in the gas analyzer for seal integrity.
Outlet pressure not the same as atmospheric pressure
Ensure that the outlet pressure is the same as the atmospheric pressure.
Conduct waste gases directly into the atmosphere or through a line with
a large internal diameter which is as short as possible, or into a gas dis-
charge line.
Do not conduct waste gases via restrictions or shut-off valves!
Measured value display is not stable
Vibrations
Provide measures to reduce the vibrations.
Increase the low pass time constant (see page 90).
Leaks in the gas feed paths
Check the gas feed paths for seal integrity (see page 117).
Loss of sensitivity
Check amplification drift display of the sample component (see page 121).
Uras26: Emitter modulation uneven
Have the emitter and modulator checked by service.