82 EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A
Measuring range configuration with ECT
Range limits
Measuring range switch-over
The type of measuring range switch-over ("Range Mode") is configured in
the "Component" dialog:
NOTE
The automatic measuring range switch-over ("Autorange") can only be con-
figured if an analog output has been assigned to the detector.
The externally controlled measuring range switch-over ("DI Control") can
only be configured if an analog output and a digital input is assigned to the
detector).
Range limits and autorange switch-over thresholds
The lower and final values of the measuring range and the autorange
changeover thresholds are configured in the measuring range dialog "Meas-
urement Range":
NOTE
The values of the autorange switch-over thresholds must both be in the
range "Measuring range 2 lower range value up to measuring range 1 final
value".
The value of the autorange switch-over threshold MR2MR1 must be less
than the value of the switch-over threshold MR1MR2.
The conditions (see page 80) for the limits and combinations of the meas-
uring ranges must be complied with.