1.3 Specifications
Power Requirements
See 1.2 Model Number Breakdown
Power Consumption
23VA typical
Refer to Converter Instruction Bulletin, as specified,
to supplement this data.
Flowmeter Characteristics:
Meter Size/Flow Capacity
See TABLE 1-2.
Span
Factory set at specified range between extremes listed in
TABLE 1-2; can be field adjusted.
Rangeability
100:1
Minimum Fluid Conductivity
5
µ
S/cm
System Accuracy:
Meter Capacity
specified on primary data tag (equal to max flow capacity
in engineering units). This factor is equivalent to a flow ve-
locity of 10 m/s in the metering tube.
Environmental Limits:
Enclosure Classification
(Standard Meter)
IEC 529 IP65; NEMA 4X, indoor or outdoor installation
Accidental Submergence Option
IEC 529 IP67; NEMA 4X, 33 ft. H
2
O/48h (10m H
2
O/48 h)
Continuous Submergence
Option
IEC 529 IP68;NEMA 4X, 33 ft. H
2
O depth (10m H
2
O depth)
Relative Humidity
10% to 90% (non-condensing)
Ambient Temperature Limits
at primary
Max:
140
o
F(60
o
C) for Process Temperatures up to 266
o
F
(130
o
C). Max. temperature is reduced by 1
o
for each de-
gree increase in process temperature above 266
o
F (130
o
C)
Min:
-13
o
F (-25
o
C)
Flowrate Accuracy
Frequency Output
<2% of meter capacity
+0.01% of meter capacity
>2% of meter capacity
+0.5% of rate
4-20 mA Output
add
+0.1% of span to above
numbers
INSTRUCTION MANUAL 10DX3111A&G Sizes 14 - 24 Inches
1-6