Navigator 600 Series Analyzers
Silica and Phosphate Appendix B – Acyclic Parameter Mapping
8 IM/NAV6PBS Issue 1
Transducer Block Parameters
Enable streams 1 145 Array Unsigned8 6 N RW
0= disabled
1 = enabled
Stream sequence 1 146 Array Unsigned8 12 N RW 1,2,3,4,5,6 1 to 6
Sample Interval 1 147 Simple Unsigned16 2 N RW 0 = continuous
Up scale sample time 1 148 Simple Unsigned8 1 N RW 10 to 30 minutes
Down scale sample time 1 149 Simple Unsigned8 1 N RW 10 to 30 minutes
Zero calibration time 1 150 Simple Unsigned16 2 N RW 10 to 60 minutes
Secondary calibration time 1 151 Simple Unsigned16 2 N RW 10 to 60 minutes
Recovery time 1 152 Simple Unsigned16 2 N RW 10 to 60 minutes
Calibration factor limit 1 153 Simple Unsigned16 2 N RW 15 to 50 (value x100) 0.15 to 0.5
Calibration offset limit 1 154 Simple Unsigned16 2 N RW
SiO2 = 50 to 500
PO4 = 2 to 200 (value x100)
P = 0 to 65 (value x100)
SiO2 = 50 to 500
PO4 = 0.02 to 20.00
P = 0.00 to 0.65
Sample pump speed 1 155 Simple Unsigned16 2 N RW 1 to 20 (value x 10) 0.1 to 2.0
Reagent pump speed 1 156 Simple Unsigned16 2 N RW 1 to 20 (value x10) 0.1 to 2.0
Sample heater proportional
band
1 157 Simple Unsigned16 2 N RW 1 to 9999 (%value x 10) 0.1 to 999.9%
Sample heater integral time 1 158 Simple Unsigned16 2 N RW 0 to 9999 seconds
Sample heater cycle time 1 159 Simple Unsigned16 2 N RW 1 to 9999 (value in seconds x 10) 0.1 to 999.9s
Reagent bottle type 1 160 Simple Unsigned8 1 N RW 0
0 = Standard
1 = Custom
Reagent bottle capacity 1 161 Simple Unsigned16 2 N RW 250 10 to 1000 litres (x 100)
0.1 to 10.0 litres – bottle type must be
'Custom' to change
Reagent limit 1 162 Simple Unsigned16 2 N RW 7 1 to 30 days
Cal solution limit 1 163 Simple Unsigned16 2 N RW 3 1 to 30 calibrations
Cleaning solution limit 1 164 Simple Unsigned16 2 N RW 3 1 to 30 cleaning cycles
1
Store definitions – refer to page 7
2
Access definitions – refer to page 7
Description Slot
Slot
Index
Object
Typ e
Data Type Bytes Store
1
Access
2
Default Valid Range Note
Table B.1 Data Structure (Sheet 2 of 38)