What to do if ABB Navigator 500 show 'Media Card Full'?
- PpmartinezSep 9, 2025
If ABB Measuring Instruments display 'Media Card Full', it means the memory card is full and no more data can be saved. Replace the memory card.

What to do if ABB Navigator 500 show 'Media Card Full'?
If ABB Measuring Instruments display 'Media Card Full', it means the memory card is full and no more data can be saved. Replace the memory card.
How to fix flow error on ABB Measuring Instruments?
If you're getting a flow error on your ABB Measuring Instruments (S1, S2, S3, S4), it means the sample flow rate is less than 50 ml/min. To resolve this, increase the sample flow to the wet-section.
What to do if ABB Navigator 500 Measuring Instruments report ADC Failure?
An ADC Failure (S1, S2, S3, S4) in ABB Measuring Instruments indicates a wet-section failure, which could be temporary or permanent, in the analog to digital converter. First, cycle power to the transmitter. If the problem continues, replace the electronics inside the wet-section.
Why did calibration fail on my ABB Navigator 500?
If the last wet-section calibration failed on your ABB Measuring Instruments, check that the reagent is entraining properly (look for bubbles at the end of the entrainment tube and ensure the reagent is fresh). Also, verify the calibration solution.
What to do if ABB Navigator 500 Measuring Instruments show 'No Sample'?
If ABB Measuring Instruments show 'No Sample (S1, S2, S3, S4)', it means there is no sample available at the wet-section. You should check that sample flow rates are greater than 50 ml/min.
What does 'No Samples' mean on ABB Measuring Instruments?
If you are using a sodium multi-stream setup with ABB Measuring Instruments, 'No Samples' means the transmitter cannot detect any samples flowing into the wet-section.
What to do if ABB Measuring Instruments show 'No Regen.Sol'?
If ABB Measuring Instruments display 'No Regen.Sol (S1, S2, S3, S4)', it indicates that you are out of regeneration solution. Replenish the solution.
What does NV Error SW Key 1 mean on ABB Navigator 500?
An NV Error SW Key 1 on ABB Measuring Instruments indicates a failure of the non-volatile memory on software key 1 board or permanent corruption of its data. Cycle power to the device. If the problem persists, check all configuration parameters and correct any errors.
What does NV Error Proc Bd mean for ABB Measuring Instruments?
An NV Error Proc Bd on ABB Measuring Instruments indicates a failure of the non-volatile memory on the processor display board or permanent corruption of its data. Start by cycling power to the transmitter. If the problem persists, check all configuration parameters and correct any errors.
What does 'Media Near Full' mean on ABB Navigator 500?
The 'Media Near Full' message on ABB Measuring Instruments means the memory card is more than 90% full. Replace the memory card to resolve this.
| Brand | ABB |
|---|---|
| Model | Navigator 500 Series |
| Category | Measuring Instruments |
| Language | English |
Be sure to read, understand and follow the instructions contained within this manual before and during use of the equipment.
Identifies potential hazards related to electrical and chemical aspects of the equipment.
Covers sample requirements, location selection, and mounting procedures for the wet-section.
Details optional accessories, location, and various mounting options for the transmitter.
Covers grounding, PCB access, and terminal connections for the wet-section.
Details power, signal, and communication wiring for the transmitter.
Details the initial setup procedures for the wet-section, including fitting electrodes and reservoir.
Describes the initial power-up, connections, and configuration of the transmitter.
Provides step-by-step instructions for performing an initial calibration of a single-stream wet-section.
Provides step-by-step instructions for performing an initial calibration of a multi-stream wet-section.
Outlines routine maintenance procedures for the wet-section, including weekly, monthly, and annual tasks.
Lists common diagnostic messages, their possible causes, and suggested remedial actions.