Navigator 500
Sodium analyzer 5 Calibration
OI/ASO550–EN Rev. G 23
5.3 Performing a calibration for the first time –
multi-stream wet-section
To perform a quick calibration from an Operator page:
1. Press the key (below the Cal prompt).
The Calibrate page is displayed:
2. Press the key (below the Select prompt).
The Calibrate / Sodium Three Stream page is displayed.
3. Press the key (below the Select prompt on the
Calibrate / Sodium Three Stream page).
The Sodium Three Stream / Start Calibration page is
displayed:
4. Press the key (below the Yes prompt on the Sodium
Three Stream / Start Calibration page).
The Calibration page is displayed with a bar graph to
indicate calibration progress:
At the end of the calibration, and if the calibration is
successful, the Active Slope, Active Offset, Last Slope and
Last Offset values in the Signals View are updated – see
Section 6.6, page 33 for details of View modes.
5. To exit the Calibration page, press the key (below the
OK prompt). Calibration continues and calibration
progress can be monitored from the Calibration View –
see Fig. 5.1, page 24.
The Calibrate / Sodium Three Stream page is displayed.
6. Press the key (below the Back prompt on the
Calibrate / Sodium Three Stream page).
The Calibrate page is displayed:
7. Press the key (below the Exit prompt) to return to the
Operator page.
CAL
2012–04–15
10 : 34: 52
Navigator
Tag 1
Tag 2
Tag 3
09:57 30/01
2.93
09:57 30/01
2.75
09:57 30/01
2.25
ppb
ppb
ppb
Calibrate
Sodium Three Stream
Back Select
Sodium Three Stream
Start Calibration
Exit Yes
Note. The Active Slope and Active Offset values are
not updated if the calibration fails. The Calibration Log
is updated with the Active Slope and Active Offset
values.
PV 2.93
SS1 100
Sodium Three Stream
Please Wait
Abort OK
ppb
ppb
Calibrate
Sodium Three Stream
Back Select