Navigator 500
Sodium analyzer 7 Menu descriptions
48 OI/ASO550–EN Rev. G
7.1.3 Device Setup
This level is used to access standard setup parameters.
Menu
Device Setup
Exit
Select
Menu Comment Default
Initial Setup
Instrument Tag A 16-character alphanumeric device (transmitter) identification tag
displayed in the title bar on all Operator pages.
Navigator
Temperature Units Selects the temperature units displayed (°C or °F). °C
Reset to Defaults Resets the configuration to the factory default settings.
Security Setup Security access levels are provided, each protected by a password of
up to 6 alphanumeric characters.
Calibrate Password Provides access to Calibrate level only.
(Not factory-set.)
Advanced Password Provides access to all configuration parameters.
(Not factory-set.)
Reset Passwords Clears all passwords.
Range Change Setup Notes.
Displayed only if 1 single-stream wet-section is connected.
Range 3 must always be the largest range, Range 1 must always
be the lowest.
Range Changing is disabled automatically when the sensor type
is changed.
Range Change When Enabled, Analogue Output 2, Relays 3 and 4 are configured
automatically for range changing and are not available via the
Input/Output sections.
Disabled
Range 1 Span
Range 2 Span
Range 3 Span
For correct operation, Range 1 Span should be the lowest range and
Range 3 Span should be the highest range.
When Enabled – the following zero and span ranges 1 to 3 are set
automatically:
Range Zero Span
1
2
3
0
0
0
100
1000
10000
The range spans can be changed to suit end-user’s requirements.
Continued on next page…